Thermo Scientific Vanquish Installation Manual page 13

Solvent monitor
Hide thumbs Also See for Vanquish:
Table of Contents

Advertisement

Solvent Monitor
Installation Guide
No.
System interlink connection between
7
Pump to autosampler
8
Autosampler to column compartment
9
Column compartment to detector
10
Detector to solvent monitor
6. If the column compartment is attached to the right side of the
Vanquish system stack and if the system stack includes a detector:
Detach the column compartment from the system stack (refer to the
Operating Manual of the column compartment).
7. If the system stack includes a detector:
a) Cut the cable cover to a length of 33 cm with a pair of scissors.
b) Insert the USB cable into the cable groove of the module nearest
to the cable.
c) Insert the system interlink cable into the same cable groove.
Figure 6: Cables in cable groove (system seen from the right side)
d) Fix the cable cover to the cable groove.
8. If the column compartment was attached to the right side of the
Vanquish system stack: Attach the column compartment to the
system stack (refer to the Operating Manual of the column
compartment).
9. If the system stack does not include a detector: Put the cables in the
gap between the column compartment and the autosampler.
3
Installation
Page 13