Thermo Scientific Vanquish Installation Manual page 22

Solvent monitor
Hide thumbs Also See for Vanquish:
Table of Contents

Advertisement

3
Installation
Page 22
7. Insert the dipstick into the waste container until the end cap is
slightly above the waste container bottom and tighten the cap onto
the waste container.
8. Press the retaining guide into the cap.
9. Adjust the dipstick so that the end cap rests on the bottom of the
waste container and that the dipstick is straight. Doing so ensures
high measurement accuracy.
10. Guide the other end of the solvent monitor line through the
detector waste outlet in the system base.
Figure 13: Guide the solvent monitor line through the detector waste
outlet in the system base (system base shown from the right side)
11. Guide the solvent monitor line from the detector waste outlet in the
system base through the tubing guides of all modules in the system
stack.
12. Guide the solvent monitor line through the guide hole in the solvent
rack.
13. Slip a line marker over the solvent monitor line to identify the
solvent monitor line at the solvent monitor.
14. If necessary, cut the solvent monitor line to length with the capillary
cutter.
15. Push the solvent monitor line straight into port 4 of the 4-channel
solvent monitor or port 8 of the 8-channel solvent monitor until you
perceive a resistance and then, push it down about 5 mm further.
16. To check proper connection, pull slightly on the line. If the line does
not slip out of the fitting, it is fixed correctly.
Solvent Monitor
Installation Guide