Thermo Scientific Vanquish Installation Manual page 21

Solvent monitor
Hide thumbs Also See for Vanquish:
Table of Contents

Advertisement

Tool required
Follow these steps
Solvent Monitor
Installation Guide
Parts not included in the ship kit:
The table shows the necessary parts and the requirements.
Needed part
Maximum
capacity
Waste
10 L
container
Capillary cutter delivered with the solvent monitor
TIP For a fast calibration, use an empty waste container.
1. Take a long solvent monitor line.
2. Optional: Cut the dipstick of the solvent monitor line with the
capillary cutter to length, so that the dipstick stands out of the waste
container by only a few centimeters, with the dipstick being straight
in the waste container.
Although this step is optional, keep in mind that a small waste
container may not be stable enough with a low waste level and a
long dipstick.
3. Slip a line label over the solvent monitor line.
4. If needed for fixation, slip a retaining guide over the dipstick.
5. Feed the dipstick of the solvent monitor line through an open hole in
a waste container cap.
6. Attach an end cap to the dipstick of the solvent monitor line.
Maximum
Remarks
height
34 cm
Further requirements:
Do not use air-tight caps, for example,
caps with air valves.
Cap with a 5.6-mm hole for fixing the
dipstick (recommended: caps from
S.C.A.T. Europe GmbH)
Fisherbrand™ waste containers (for
ordering information, contact your
local Thermo Fisher Scientific sales
organization)
Constant cross-section for most of the
waste container height
Recommendations:
Maximum distance between waste
container and solvent monitor: 3 m.
For the first operation, use an empty
waste container.
3
Installation
Page 21