3. Optional: Cut the dipstick of the solvent monitor line to length with
the capillary cutter, so that the dipstick stands out of the solvent
reservoir by only a few centimeters, with the dipstick being straight
in the solvent reservoir. Although this step is optional, keep in mind
that a small solvent reservoir may tilt with a low solvent level and a
long dipstick.
4
Figure 8: Cutting the dipstick to length
No.
Description
1
Solvent line
2
Solvent monitoring line
3
Location to cut the dipstick to length with the capillary cutter
4
Dipstick
4. Slip a line label over the solvent monitor line.
5. Slip a retaining guide over the dipstick.
6. Feed the dipstick of the solvent monitor line through an open hole in
a cap.
7. Attach an end cap to the dipstick of the solvent monitor line.
8. Insert the dipstick into the solvent reservoir until the end cap is
slightly above the solvent reservoir bottom and tighten the cap onto
the solvent reservoir.
9. Press the retaining guide into the cap.