Thermo Scientific Vanquish Installation Manual page 18

Solvent monitor
Hide thumbs Also See for Vanquish:
Table of Contents

Advertisement

3
Installation
Page 18
12. Cut the solvent monitor line to length with the capillary cutter.
Figure 10: Cutting the solvent monitor line to length
No.
Description
1
Solvent line to the pump
2
Solvent monitor line
3
Location to cut the solvent monitor line to length with the capillary cutter
13. Slip an appropriate second line label over the other end of the
solvent monitor line to identify the solvent monitor line at the
solvent monitor.
14. Insert the solvent monitor lines into the solvent monitor:
15. Push the solvent monitor line straight into one of the ports of the
solvent monitor until you perceive a resistance and then, push it
down about 5 mm further. The ports to use depend on the solvent
monitor type and the pump type.
The first table shows the assignment for the 4-channel version and
the second table shows the assignment for the 8-channel version.
Pump type
Port 1
Quaternary
A
Binary
A1
Dual
A
Isocratic
A
Port 2
B
A2
B
L
L
-
1
2
3
Port 3
Port 4
C
Waste
B1
B2
A
B
R
R
-
Waste