Troubleshooting - Thermo Scientific Shandon TissueWave 2 Operator's Manual

Microwave specimen processing system
Table of Contents

Advertisement

Warning Screens
42
TissueWave 2 Operator Guide

5 - Troubleshooting

The Shandon TissueWave 2 provides informative
Audible Alert
and an
The following table lists the
possible remedies for the appropriate faults.
Error Message
when certain conditions or problems occur.
Warning Screens
Cause
Instrument is in the process
of powering up
Airflow has been
significantly reduced
A key has been pressed for
more than 3 seconds (See
Stuck Key Identification)
Measured temperature has
exceeded the set-point
temperature by 6°C
Warning Screens
, their cause(s) and
Remedy
• None - This is normal
Check for obstructions in the air-flow
circuit
Check for correct venting system
Check that the hose is properly attached
Turn power off and then back on
If problems persist call ThermoFisher
Scientific Technical Support
This is a fatal error
Turn power off and then back on
If problems persist call ThermoFisher
Scientific Technical Support
Check fluid level is appropriate
Check power settings
Check temperature probe placement and
connections
Check probe for damage
If problems persist call ThermoFisher
Scientific Technical Support