Warning Screens
42
TissueWave 2 Operator Guide
5 - Troubleshooting
The Shandon TissueWave 2 provides informative
Audible Alert
and an
The following table lists the
possible remedies for the appropriate faults.
Error Message
when certain conditions or problems occur.
Warning Screens
Cause
Instrument is in the process
of powering up
Airflow has been
significantly reduced
A key has been pressed for
more than 3 seconds (See
Stuck Key Identification)
Measured temperature has
exceeded the set-point
temperature by 6°C
Warning Screens
, their cause(s) and
Remedy
• None - This is normal
Check for obstructions in the air-flow
•
circuit
•
Check for correct venting system
•
Check that the hose is properly attached
•
Turn power off and then back on
If problems persist call ThermoFisher
•
Scientific Technical Support
•
This is a fatal error
•
Turn power off and then back on
If problems persist call ThermoFisher
•
Scientific Technical Support
•
Check fluid level is appropriate
•
Check power settings
Check temperature probe placement and
•
connections
•
Check probe for damage
If problems persist call ThermoFisher
•
Scientific Technical Support