6
TissueWave 2 Operator Guide
The Shandon TissueWave 2 is designed and made with care.
The instrument is safe to use, simple to operate, and easy to maintain.
Many safety features have been built into the unit to ensure its safe
operation.
However, incorrect actions by a user may damage the equipment, or
cause a hazard to health.
All users must read and understand the following paragraphs
! !
before using this instrument.
Operate the unit from a properly grounded and
! !
appropriately rated receptacle.
To minimize the risk of electrical shock and the possibility
of interference with other electronic devices, the unit must
always be powered from a grounded receptacle.
Position the Shandon TissueWave 2 so that it is possible to
interrupt the mains supply at the source by removing the plug
from the socket.
In addition, the unit should be operated on its own electrical
power circuit with a current rating at least equal to the unit's
fuse rating. Line voltage may fluctuate particularly if other
devices are powered on the same line.
The shortest possible distance between the service outlet
and the main connection feeding the service will provide
the best results in terms of consistent power output from the
microwave.
! !
Ensure proper operation of the vent system.

Safety Information