Typical Loads - W&H Lexa PLUS Instructions For Use Manual

Table of Contents

Advertisement

Load type
Placement
On trays allowing adequate space in-between
Pouched items
bags. Ensure that packs do not touch the walls of
the chamber. Place sterilization pouched items
with the paper side facing up.
Items made from different
On separate trays or wrapped/pouched
materials (stainless steel,
carbon steel, aluminium, etc.)
Instruments manufactured
Place paper among them and the trays to avoid
from carbon steel
rusty spots
Lexa PLUS  | Instructions for Use |  RIS-311 ENG Rev03 |  29/08/2022 |  © W&H Sterilization Srl
PARTIAL LOAD
If the chamber is just partially loaded, place the load in such a way
that the space in-between the trays is maximized. Spread items
evenly on multiple trays. Below is a example with three trays.

Typical loads

EXAMPLES
An example of a typical load used during the tests performed by
W&H is reported below:
 
Sterilization cycles
51

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ris-311

Table of Contents