Extraordinary maintenance

DRAINING THE USED AND CLEAN WATER TANK
If you left accidentally the tanks full for more than seven days or if you plan not to use the sterilizer for at least seven days, you have to drain the
tanks.
 
Open the sterilizer service door.
1
 
Put a container below the sterilizer (1.45 gal (5.5 l) minimum) and place the end
2
of the drain tube in it.
 
To drain the used water, insert the drain tube connector in the grey connection.
3
 
To drain the clean water, insert the drain tube connector in the blue connection.
4
 
When the water has been completely drained, press the release button to remove
5
the drain tube and close the service door.
Lexa PLUS  | Instructions for Use |  RIS-311 ENG Rev03 |  29/08/2022 |  © W&H Sterilization Srl