GENERAL RECOMMENDATIONS
Follow these recommendations to obtain the most from the drying:
 
Ensure the paper side of the sterilization bags faces up, and
n
that the space in-between bags is enough.
 
To enjoy the full benefit of short cycle times when only one
n
tray is used, always place the load on the upper tray of the
chamber rack and remove all other trays from the chamber.

Sterilization cycle description

AVAILABLE STERILIZATION  CYCLES
See "Sterilization programs" on page  90 for the full list of key
program features, including sterilization time, temperature and
recommended load type.
AVAILABLE ECO DRY MODES
Icon
Mode
ECO DRY
The drying time is automatically adjusted to the shortest
plus
possible for the total amount of load:  the smaller the load,
the faster the cycle. It ensures perfect drying of standard
loads.
ECO DRY The drying time is automatically adjusted to the total
amount of load. It is longer than the ECO DRY plus mode.
It is useful with those particular loads that require an
extended drying time (suggested for container load).
Lexa PLUS  | Instructions for Use |  RIS-311 ENG Rev03 |  29/08/2022 |  © W&H Sterilization Srl
Description
Icon
Mode
ECO DRY
The drying time is automatically adjusted to the total
+25%
amount of load with a 25% of additional drying duration.
ECO DRY
The drying time is adjusted to the total amount of load with
+50%
a 50% of additional drying duration.
Sterilization cycles
Description
53