Maintenance
Monthly or 50-cycle maintenance
CLEANING THE DOOR GASKET AND THE CHAMBER FACE SIDE
Proceed as follows:
Clean the door gasket and the outer edge of the chamber with a non-abrasive
1
cloth moistened with clean water.
Notice:
Do not use abrasive products, cutting tools or sharp objects.
Rinse with clean water.
1
Note: when the seal is new it might be necessary to hold the door gently closed at the
sterilization start.
66
Lexa PLUS | Instructions for Use | RIS-311 ENG Rev03 | 29/08/2022 | © W&H Sterilization Srl