10 ERROR CODES

Press
to clear error codes.
Error code Explanation
Unable to obtain a stable reading. This is either due to a very high electrical noise
0 Err
environment, beyond emc specification limits, or could indicate a gauge fault.
Calibration factors are out of range. Clear the error message and repeat the adjustment
1
Err
procedure.
Averaging error.
2 Err
No readings taken for average calculation.
3 Err
Statistics memory full.
4
Err
Statistics overflow, more than 9,999 readings have been taken. Statistics data must be
5
Err
cleared (see Section 6.3) before further readings can be taken for statistical analysis.
6 Err
Gauge over-range reading.
Memory hardware fault. Contact Elcometer or your local Elcometer supplier.
7
Err
Gauge is not configured after repair. Contact Elcometer or your local Elcometer supplier.
8 Err
The printer is off line, out of paper or printer or cable fault. Correct printer fault or press
9
Err
to cancel print function.