The Correction Value Method - Elcometer 345 Operating Instructions Manual

Coating thickness gauge
Hide thumbs Also See for 345:
Table of Contents

Advertisement

4. Press and hold
5. Press
to accept the low calibration or wait 7 seconds to accept the calibration automatically.
The display will now show CH1 (= Calibrate High).
6. Place the chosen high calibration foil on the bare metal surface and take a number of readings until a
stable average is displayed.
7. Press and hold
8. Press
to accept the calibration or wait 7 seconds for the gauge to accept the calibration
automatically. The display will blink and then re-display the reading. The bleeper will sound 3 times to
indicate that this routine is complete.
9. Confirm that the average gauge reading on each calibration foil is correct. An intermediate foil
thickness can also be used to confirm the performance over the calibrated range.
The calibration values are stored in the gauge until new values are entered, even when the gauge is
switched off.
on the display indicates the special calibration on switch on.
To re-calibrate, repeat the procedure. Re-calibrating using the smooth surface calibration method (see
"Calibration Adjustment for Smooth Surfaces" on page 17) to set zero will remove

5.4 THE CORRECTION VALUE METHOD

The smooth surface adjustment method (see "Calibration Adjustment for Smooth Surfaces" on page 17)
can be used for blast cleaned surfaces if a correction value is applied to each reading taken on the
roughened surface.
or
for more than 5 seconds to adjust this average to the foil value.
or
for more than 4 seconds to adjust this average to the foil value.
will stop flashing and will remain visible on the display.
R
from the display.
19