iTero Element 2 and iTero Element Flex User Manual
Test
Emission (per ETSI EN 301 489-1)
Relevant for iTero Element 2 only
Conducted emissions on
mains terminals in freq.
range 150 kHz - 30 MHz
Radiated emissions
in freq. range 30 - 6000 MHz
Harmonic current test
Flicker tests
Immunity (per ETSI EN 301 489-1)
Relevant for iTero Element 2 only
Immunity from Electrostatic
discharge (ESD)
Immunity from radiated
electromagnetic fields
Immunity from Electrical
Fast transient (EFT)
Immunity from Surge
Immunity from conducted
disturbances induced by
radio-frequency fields
Immunity from Voltage
interruptions
© 2019 Align Technology, Inc. All rights reserved.
Standard
Severity level/lines
ETSI EN 301 489-
Group 1 Class B
1;
230 VAC mains
EN 55032
ETSI EN 301 489-
Class B
1;
EN 55032
ETSI EN 301 489-
230 VAC mains
1;
EN 61000-3-2
ETSI EN 301 489-
230 VAC mains
1
EN 61000-3-3
EN 61000-4-2
4 kV contact discharge
8 kV air discharge
EN 61000-4-3
3.0 V/m,
80 MHz 6.0 GHz,
80% AM, 1 kHz
EN 61000-4-4
AC mains: ± 1.0 kV;
Tr/Th – 5/50 ns, 5 kHz
EN 61000-4-5
AC mains: ± 1.0 kV DM / ± 2.0 kV CM,
Tr/Th – 1.2/50 (8/20) s
EN 61000-4-6
AC mains: 3.0 V
0.15÷80 MHz, 80% AM @ 1 kHz
EN 61000-4-11
AC mains:
0 % - 0.5 cycle & 1 cycle;
70% - 25 cycles; 0% - 250 cycles
;
RMS
Test
results
Complies
Complies
Complies
Complies
Complies
Complies
Complies
Complies
Complies
Complies
103
Need help?
Do you have a question about the Element 2 and is the answer not in the manual?
Questions and answers