iTero Element 2 and iTero Element Flex User Manual
Safety instructions
Before beginning to work with the system, all users are required to read these safety instructions.
Power supply
Battery power
Electric
warnings
Wireless LAN
Safety
classifications
Prescription
health device
viii
Power is supplied to the system via an internal medical-grade power supply.
iTero Element 2 only:
Charging – the scanner battery will be fully charged after being plugged into a
power source for 2 hours.
With a fully-charged battery, you can scan up to 30 minutes.
Warning: The scanner is provided with a Li-ion rechargeable battery pack. There
is a danger of explosion if the battery is incorrectly replaced. Replace only with
the same type recommended by the manufacturer. Discard used batteries
according to the manufacturer's instructions.
Electric shock hazard!! Only authorized Align Technology technicians can
remove external panels and covers. There are no user-serviceable parts inside.
To avoid the risk of electric shock, the scanner must be connected only to a
mains supply with protective grounding.
Only an Align Technology approved web camera or disk on key should be
connected to the USB sockets on the rear of the touch screen.
iTero Element Flex only:
iTero Element Flex is provided with a hub containing a power supply for the
wand. The Flex hub must be kept dry and be protected against breakage.
Connect the Flex hub only to a laptop that is approved according to IEC60950
and to UL60950-1. The laptop and all its accessories should be located at least
1.5m away from the patient. Do not scan a patient and touch the laptop or any of
its accessories at the same time.
Only the iTero wand and the approved laptop should be connected to the USB
sockets on the Flex hub.
Only the Align Technology approved power cable should be used to connect the
Flex hub to the AC outlet.
The system comes equipped with a wireless LAN unit.
Type of protection against electrical shock: Class 1.
Degree of protection against electrical shock: Type BF.
Degree of protection against harmful ingress of water: Ordinary.
Equipment not suitable for use in the presence of flammable anesthetic mixtures.
Mode of operation: Continuous.
The system serves as a prescription medical device and should be operated by
qualified health-care providers only.
iTero Element 2 and iTero Element Flex User Manual
© 2019 Align Technology, Inc. All rights reserved.