Inspection Before Use; Inspection For Appropriate Connection; Inspection For Material Surface - Hitachi EUP-CC531S Instruction Manual

Biplane transrectal / vaginal probe
Table of Contents

Advertisement

2

Inspection before Use

Prior to use, the probe and accessories must be carefully inspected
that they are appropriate for use. If you find any damage, do not
use them and contact a service support immediately.
2.1

Inspection for Appropriate Connection

2.1.1 Confirm that the system is correctly operating. Refer to
the instruction manual for the ultrasound diagnostic
scanner.
2.1.2 Do not attach or connect unauthorized devices or
instruments on the probe, such as unauthorized biopsy
attachments.
2.1.3 Confirm that the Sterile Puncture Adapter, Puncture Guide
Fixture and software version and then settings of the
scanner are appropriate for the probe. Attach the Sterile
Puncture Adapter or Puncture Guide Fixture on the probe.
Set the main unit to display the "Needle guide line". (Refer
to the operation manual for the main unit.) Keep the probe
head in the water and insert a puncture needle in the Sterile
Puncture Adapter or Puncture Guide Fixture. Then, confirm
that the needle is inserted smoothly and the echo of the
needle is displayed on the dot line "Needle guide line"
on the monitor.
2.2

Inspection for Material Surface

2.2.1 Visually inspect the surface of the probe head, housing
and cable for any crack, scratch or denaturalization.
2.2.2 Visually inspect the envelope of the Sterile Puncture
Adapter or Puncture Guide Fixture for any break,
deformation, crack or denaturalization.
2.2.3 Visually inspect the Magnetic sensor attachment and the
Spacer for EZU-RV2S for any crack, deformation or
denaturalization.
2.2.4 Confirm whether there is the Spacer for EZU-RV2S nearby
when use RVS with EZU-RV2S.
The Spacer for EZU-RV2S is small, please do not lose the Spacer.