Biplane transrectal / vaginal probe
Table of Contents

Advertisement

c) Applicable brush, e.g. REF 09098 (Interlock)
d) Two tanks, one for cleaning and one for disinfection - optional:
1 additional tank for rinsing with deionized/tap water
e) 50 ml syringe
f) Soft, fluff free cloth or single use towel
g) Personal protective equipment (gloves, water repellent protective
skirt, face protection mask or protective glasses, see also
instructions of the manufacturer for the detergent and the
disinfectant)
Manual Cleaning:
Prepare the detergent solution in a tank with cold water (please follow
the instructions of the detergent manufacturer regarding application,
dilution and contact time).
1) The temperature of the detergent solution should be between 15-30 °C,
concentration is 1.6%. Please note the minimum contact time of the
detergent in the manufacturer's instruction. If a differing
detergent is used, please also note the approved material
compatibility for the medical device.
2) Immerse the parts of the EZU-PA3U into the diluted detergent solution.
Clean them under the surface of the detergent solution with a brush
to remove all visible soil. Brush the whole length of the cavity
at least 5x using an applicable brush.
3) Using a syringe flush the cavity of the puncture guide fixture 5x
under the surface of the detergent solution with 50 ml diluted
detergent.
4) Wipe the parts of the puncture guide fixture under the surface of
the detergent solution with a soft, fluff free cloth. Be sure that
all grooves of the puncture guide fixture are implemented during
the cleaning process.
5) The parts of the EZU-PA3U should be left in the detergent solution
according to the specified contact time of the detergent
manufacturer.
6) Rinse the parts of the EZU-PA3U with running tap water for 1 minute.
- 38 -
Q1E-EP1357