Cleaning, Disinfection And Sterilization Of Ezu-Pa3U - Hitachi EUP-CC531S Instruction Manual

Biplane transrectal / vaginal probe
Table of Contents

Advertisement

6 Cleaning, Disinfection and Sterilization of EZU-PA3U

EZU-PA3U must be reprocessed after each use. Refer to the
reprocessing instruction in this chapter.
WARNINGS
Transportation
before using
Levels of reprocessing requirements:
Refer to the chapter 5.
EZU-PA3U is delivered unsterile. Prior to the first
use, reprocess it.
The cavities of the puncture guide fixture require
particular attentions during all processes. The
puncture guide fixture must be dismantled into single
parts before reprocessing.
EZU-PA3U should be packed in a sterile pouch or
container to transport from Central Sterile Supply
Department (CSSD) to an operating room. Be careful
not to damage the sterile pouch or container during
transportation.
- 35 -
Q1E-EP1357

Advertisement

Table of Contents
loading

Table of Contents