6.1

Point of use (Pre-cleaning)

Pre-cleaning should be done immediately after each
use. The procedure is as follows:
1) Remove the EZU-PA3U from the probe.
2) Flush patient's blood or fluid by tap water from the puncture guide
fixture directly after use, until the surface looks visually clean.
3) Disassemble the EZU-PA3U (Fig.21) and immerse all parts in
sufficient amount of high quality tap water. Scrub them using soft
cloth to remove all visible soil and dried protein from their
surface.
6.2

Containment and transportation

Refer to 5.2.
6.3

Manual Cleaning and disinfection

Prepare
following
cleaning and disinfection:
a) Detergent: Cidezyme (Johnson & Johnson,
#2258) or another cleaning agent with
approved material compatibility for
this medical device.
b) Disinfectant: Cidex OPA (Johnson &
Johnson, # 20391) or another
disinfectant with approved material
compatibility for this medical device.
Fig.21 Disassembly of EZU-PA3U
items
before
- 37 -
Manual cleaning and
manual
Rinsing after manual
Manual Disinfection
Rinsing after manual
Point of use
(Pre-cleaning)
disinfection
Manual Cleaning
cleaning
disinfection
Q1E-EP1357