Inspection Before Use; Inspection For Appropriate Connection; Inspection For Material Surface - Hitachi EUP-V73W Instruction Manual

Transvaginal probe
Table of Contents

Advertisement

 
2
Inspection before Use 
Prior to use, the probe and accessories must be carefully inspected for appropriate use. If 
you find any damage, do not use them and contact a service support immediately. 
 
2.1
Inspection for Appropriate Connection 
1) Confirm that the ultrasound diagnostic scanner is correctly operating.    Please refer to 
the instruction manual of the scanner for the operation. 
2) Confirm  that  any  unauthorized  device/instrument  such  as  an  unauthorized  biopsy 
attachment is not attached or connected to the probe.   
3) Confirm that the Sterile Puncture Adapter and settings of the scanner are appropriate 
for  the  probe.  Attach  the  Sterile  Puncture  Adapter  to  the  probe  according  to  "3. 
Operation Procedure". Display the "Needle guide line" on the monitor. (Refer to the 
operation  manual  of  the  ultrasound  diagnostic  scanner.)  Keep  the  probe  head  under 
sterile  water  and  insert  a  puncture  needle  in  the  Sterile  Puncture  Adapter,  then 
confirm that the needle is inserted smoothly and the echo of the needle is displayed on 
the "Needle guide line"(dot line) on the monitor. 
 
2.2
Inspection for Material Surface 
1) Visually inspect the surface of the probe head, housing and cable for any crack, scratch 
or denaturalization. 
2) Visually  inspect  the  envelope  of  the  Sterile  Puncture  Adapter  for  any  break, 
deformation, crack or denaturalization. If you find any damage, do not use the Sterile 
Puncture Adapter. 
3) Visually inspect the Magnetic sensor attachment and the Spacer for EZU‐RV2S for any 
crack, deformation or denaturalization.   
 
 
Beware not to lose the Spacer for EZU‐RV2S since it is so small. 
 
 
 
CAUTION
 
‐4‐ 
Q1E‐EP1430