Hitachi EUP-CC531S Instruction Manual page 34

Biplane transrectal / vaginal probe
Table of Contents

Advertisement

Manual Cleaning:
Prepare the detergent solution in a tank with cold water (please follow
the instructions of the detergent manufacturer regarding application,
dilution and contact time).
A) EUP-CC531S
1)
The temperature of the detergent solution should be between 15-30 °C,
concentration is 1.6%. Please note the minimum contact time of the
detergent in the manufacturer's instruction. If a differing
detergent is used, please also note the approved material
compatibility for the medical device.
2)
Immerse the immersible part of the probe without connector into
the diluted detergent solution. Wipe the immersible part of the
probe under the surface of the detergent solution with a soft cloth
to remove all visible soil.
3)
Use the syringe via the Luer inlet to fill and rinse the lumen with
the detergent solution. The lumen of the probe is rinsed 5 times
under the surface of the detergent solution with 50 ml diluted
detergent. Drain the detergent out of the lumen by using the syringe.
Afterwards, remove the syringe from the probe.
4)
Insert the cleaning brush into the lumen via the Luer inlet until
the end of the channel is reached. Brush the whole length of the
lumen 5 or more times. In addition all immersible parts of the probe
are wiped with a soft cloth until visually clean.
The inside of the water channel is very sensitive, so if the water channel
is cleaned, cleaning brush should be operated carefully. If brushing
is rough, water channel may suffer serious damage.
5)
Rinse the lumen 5 times with 50 ml detergent solution by using the
syringe.
6)
The immersible part of the probe should be left in the detergent
solution according to the specified contact time of the detergent
manufacturer.
7)
Wipe the un-immersible parts of the probe with a soft cloth dipped
with the detergent solution.
CAUTION
- 27 -
Q1E-EP1357