Automated Cleaning And Disinfection - Hitachi EUP-CC531S Instruction Manual

Biplane transrectal / vaginal probe
Table of Contents

Advertisement

(alternatively: immerse them in a tray filled with deionized
water/tap water (see Fig.17) for 5 min. and rinse the cavity of the
puncture guide fixture with 50 ml tap water. Repeat this 5x.)
7) Visually check the outer surface of the parts of the EZU-PA3U for
cleanness. If necessary, use magnifying glass for visually check.
If there is still soil visible, repeat all above steps.
Manual disinfection:
1) Prepare the disinfectant solution as stated in the procedure for
the probe.
2) Immerse the parts of the EZU-PA3U into the disinfectant. Rinse the
cavity of the puncture guide fixture with 50 ml disinfectant solution.
Repeat this 4x. Set a clock to insure the recommended contact time
which is 5 minutes.
3) Rinse the parts of the EZU-PA3U with deionized water for 1 minute.
(alternatively: immerse them in a tray filled with deionized water
for 5 min. and rinse the cavity of the puncture guide fixture with
50 ml deionized water. Repeat this 5x.)
4) Visually check the outer surface of the parts of the EZU-PA3U for
leavings of the disinfectant. If necessary, repeat the rinsing.
6.4

Automated cleaning and disinfection

The following items must be provided prior
to automated cleaning and disinfection:
a) Washer disinfector: according to DIN
EN ISO 15883 with chemo-thermal
program.
b) Detergent: Korsolex Endo-Cleaner (Bode Chemie; # 972 020)
c) Disinfectant: Korsolex Endo Disinfectant (Bode Chemie; # 972 030)
d) Washer disinfector accessories:
Basket for holding the puncture guide fixture
Basket with lid for holding the dismantled small parts (e.g. screws)
of the puncture guide fixture
Washer Disinfector
Automated Cleaning
Automated Disinfection
- 39 -
Q1E-EP1357