Containment And Transportation; Manual Cleaning And Disinfection - Hitachi EUP-CC531S Instruction Manual

Biplane transrectal / vaginal probe
Table of Contents

Advertisement

5.2

Containment and transportation

Putting the contaminated equipment into exclusive
shock and damage proof container for transportation
is recommended. It is recommended that instruments
are reprocessed as soon as possible and not later
than 4 hours after usage.
5.3

Manual Cleaning and disinfection

Prepare
following
cleaning and disinfection:
a) Detergent: Cidezyme® or another cleaning
agent with approved material
compatibility for this medical device
(Johnson & Johnson, #2258)
b) Disinfectant: Cidex® OPA or another
disinfectant with approved material
compatibility for this medical device
(Johnson & Johnson, # 20391)
c)
Cleaning brushes if applicable, i.e. Olympus Model Nr. BW-400L
(cleaning the lumen of the water channel)
d) Two tanks, one for cleaning and one for disinfection - optional:
- 1 additional tank for rinsing with deionized/tap water (sufficient
size for immersion of the immersible part of the probe at full length)
e) Syringe 50 ml with Luer-lock
f) Soft, fluff free cloth or single use towel
g) Personal protective equipment (gloves, water repellent protective
skirt, face protection mask or protective glasses, see also
instructions of the manufacturer for the detergent and the
disinfectant)
items
before
- 26 -
Manual cleaning and
disinfection
manual
Manual Cleaning
Rinsing after manual
Manual Disinfection
Rinsing after manual
disinfection
Containment and
transportation
cleaning
Q1E-EP1357