5.1

Point of use (Pre-cleaning)

Pre-cleaning should be done immediately after each
use. The procedure is as follows:
A) EUP-CC531S
1) Remove the protective cover.
2) Flush the lumen of the probe directly after use with 50 ml deionized
water/tap water using a 50 ml syringe (adaptation of the syringe
to the Luer inlet of the probe).
3) Clean the probe of all patient's blood or fluid with running tap
water until the surface of the probe looks visually clean.
4) Wipe the whole surface of the probe with gauze pad and remove
superficial visible impurities.
B) Magnetic sensor attachment and Spacer for EZU-RV2S
1) Remove the Magnetic sensor attachment and the magnetic sensor from
the probe. Remove the Spacer for EZU-RV2S from the magnetic sensor
if the sensor is EZU-RV2S.
2) Immerse the Magnetic sensor attachment and the Spacer for EZU-RV2S
in sufficient amount of high quality tap water. Scrub them using
soft cloth to remove all visible soil from their surface.
- 25 -
Point of use
(Pre-cleaning)
Q1E-EP1357