Symptom
Mattress appears to
be bottoming out.
Blower Unit Sounds
noisier than usual.
Blower motor inside
pump sounds
excessively noisy
during continuous
use.
Mattress appears to
ignore therapy
settings
Possible Cause
Comfort Control
Assess comfort control
Settings too soft.
settings.
Patient Weight.
Check patient weight is within
limit for system.
Air leak from
Check connections.
manifold.
Loose hose
Check connections in tubeset
connector.
connector.
Torn or damaged
Check.
cell.
The blower
Check to ensure the mattress
inside the pump
is fully inflated.
unit may be
approaching the
end of its life.
Pump is not
Check.
responding to
therapy settings.
Check
15
Action
Adjust as necessary.
Refer to User Manual.
Page 4 Indications.
Remove connection, and
reconnect.
Reconnect if possible,
otherwise call Service
Engineer.
Call service engineer to
replace damaged cell.
Call service engineer to
request a replacement pump
unit (if rental system).
Reset Pump and re-enter
therapy settings.