Thermo Scientific Nicolet iS50 Getting Started page 5

Spectrometer
Hide thumbs Also See for Nicolet iS50:

Advertisement

B
Humidity indicator
C
Desiccant access
cover
D
Power switch
The humidity indicator monitors the level of humidity inside the
spectrometer. The status is conveyed by the color of the
indicator:
Your spectrometer is protected from excessive humidity by
desiccant located in a compartment below the desiccant access
cover. Use the humidity indicator (see below) to monitor the
moisture level inside the spectrometer and replace the desiccant
as soon as it becomes saturated. For information about
replacing the desiccant, find "Replacing the Desiccant" in the
spectrometer Help.
Leave the desiccant in the spectrometer even if you purchased
the optional purge kit so the system is protected in the event of a
power failure or if the purge is turned off. For information about
purging your spectrometer, refer to your spectrometer Site and
Safety Guide.
The power switch turns the spectrometer on and off. Two
indicator lights next to the power switch tell you the status of the
spectrometer:
System status (steady blue - ready; flashing yellow - error with a
component).
System scan (blinking blue - scanning; off - not scanning or
sleep mode).