Thermo Scientific Nicolet iS50 Getting Started page 13

Spectrometer
Hide thumbs Also See for Nicolet iS50:

Advertisement

If you have ...
The standard sample holder
and baseplate (included with
the iS50 spectrometer)
The iS50 ATR module or an
ATR sample compartment
accessory
A DRIFTS accessory
A Specular Reflection
accessory, such as the Smart
SAGA accessory
You Should ...
Install the baseplate by placing it in the sample
compartment over the alignment pins.
Make sure no samples that would block the beam are
installed in the accessory.
Make sure the pressure device is not in contact with the
crystal. Clean the crystal with a clean, soft cloth or a white
paper napkin. If needed, you can dampen the cloth or
napkin with water, mild soap, or alcohol for the iS50 ATR
module. (If you are using a sample compartment ATR
accessory, check the documentation that came with it for
cleaning agents that are compatible with your accessory
and the type of crystal it has.)
Do not pour liquids directly onto the sampling area; if
additional solvent is needed to clean the crystal, use a
pipette or dropper.
Make sure the slider is pushed in to the second cup (the
one closer to the handle). This positions the gold alignment
mirror in the beam.
Place a gold mirror face down on the accessory.
13
Nicolet iS50 SpectrometerGetting Started
Collecting a Spectrum