11.2 Test Solutions

90
9. All signals must be stable and changing fast (within seconds).
10. Use the same process to test Skimmer bias with full He gas flow 200ml/min. Make
sure the results similar to the Figure 11-4.
Print the pages for H
Figure 11-3
"BOOST" operation in iCRC mode, maximum H
Figure 11-4
"BOOST" operation in iCRC mode, maximum He gas flow (200ml/min)
The standard installation performance tests and calibrations require a blank solution
and two multi-element solutions.
Multi-Element AJ Tuning Solution: The solutions consist of
 50 ppb of Be, Mg, Co, In, Ba, Ce, Tl, Pb, Th.
 1 ppb of Be, Mg, Co, In, Ba, Ce, Tl, Pb, Th.
The primary source of these elements is the nitrate.
and He BOOST test (Figure 11-3 and Figure 11-4).