Performance
11.2 Test Solutions
90
9. All signals must be stable and changing fast (within seconds).
10. Use the same process to test Skimmer bias with full He gas flow 200ml/min. Make
sure the results similar to the Figure 11-4.
Print the pages for H
Figure 11-3
"BOOST" operation in iCRC mode, maximum H
Figure 11-4
"BOOST" operation in iCRC mode, maximum He gas flow (200ml/min)
The standard installation performance tests and calibrations require a blank solution
and two multi-element solutions.
Multi-Element AJ Tuning Solution: The solutions consist of
50 ppb of Be, Mg, Co, In, Ba, Ce, Tl, Pb, Th.
1 ppb of Be, Mg, Co, In, Ba, Ce, Tl, Pb, Th.
The primary source of these elements is the nitrate.
and He BOOST test (Figure 11-3 and Figure 11-4).
2
PlasmaQuant MS Series
gas flow (200 ml/min)
2