Exposure Time Verification and kV-Verification

(conventional radiography technique)
For testing the exposure times a Digital Dosimeter is required (PTW Diados).
Place the Dose Detector in a distance of 8" in front of the Tube Head.
Switch unit ON.
CAUTION RADIATION! Make an exposure with:
Voltage:
Exposure time:
The reading should be 0.4 sec ±0.04 sec.
kV = 60kV ±6kV. Record reading
Switch unit OFF
If the measured Radiation Time is not within specified tolerance,
see Service Manual, chapter "Exposure Time Verification".
If the measured kV Verification is not within specified tolerance,
see Service Manual, chapter "kV Verification".