General info
Service
Read the following carefully, for information on
technical safety and responsibility on
Gram BioLine products.
WARNING –
DO NOT OPEN, MAINTAIN OR SERVICE
IN AN AREA WHEN AN EXPLOSIVE
ATMOSPHERE IS PRESENT
When servicing –
Make sure the appliance is switched Off at
the socket before service is performed on
the cabinet.
It is not sufficient to switch Off the cabinet
on the On/Off
persist in some electrical parts of the
cabinet.
Warranty may be void in the instance the cabinet is
being used for applications other than its intended
use, or otherwise not in accordance with the
guidelines stipulated in the instructions for use.
Defective parts must be replaced with original
parts from Gram BioLine. Gram BioLine can only
guarantee functional and safety requirements on
the cabinets, if above mentioned is adhered to.
The cabinet should be checked at least once yearly
by a Gram BioLine-authorized technician.
The refrigeration system and the hermetically
sealed compressor require no maintenance.
However the condenser requires regular cleaning.
Be aware that cabinets using hydrocarbons (HC)
as refrigerant, may require special handling by
qualified technicians.
44
button, as current will
Type/number plate
If refrigeration fails, first look to see whether the cabinet has been
unintentionally switched off, or whether a fuse has blown.
If the cause of failure cannot be found, contact your supplier
quoting Type and S/N. This information can be found on the type/
number plate.