Quick guide – BioCompact II
Parameter settings
On/Off
Press the
button to turn the cabinet on. Press the
cabinet will be shown when turning the cabinet on, followed by the software variant. The cabinet is ready when the temperature is
displayed. The cabinet will automatically start a defrost-cycle when turned on, and terminate it again after a system check.
Setting the temperature
Temperature adjustments are done by holding the
Confirm the settings by letting go of the buttons.
Lighting inside the cabinet – Only applicable to BioCompact ll cabinets with glass door
The settings for the lighting inside the cabinet can be changed by pressing and holding
There are two settings:
1) Light turns on when the door is open (off when closed)
2) Light is always on
User menu and alarm settings
Menu Access
+
Local alarm settings
Offset of sensors
Low-temperature protection
2
N/A
Defrosting
Alarms A2, A3, A4, A5
button for 6 seconds to turn the cabinet off. The software version of the
button and pressing either
Display code and its message
LAL
LHL
[° C]
Upper temperature limit. Code for activated alarm [A2]
LLL
[° C]
Lower temperature limit. Code for activated alarm [A3]
LHd
[min.]
Delay of upper temperature limit
LLd
[min.]
Delay of lower temperature limit
dA
On/off Door alarm. Code for activated alarm [ A1]. [1=on/0=off]
dAd
[min.]
Delay of door alarm
bU
On/off Acoustic signal for alarm codes [ A1], [ A2] and [ A3]. [1=on/0=off]
CAL
CA
[K]
Offset setting for A-sensor. Reference sensor for the refrigeration system
CE
[K]
Offset setting for E-sensor. Reference sensor for the display and alarms
Offset setting for F-sensor. Reference sensor for the low-temperature
CF
[K]
protection
FP
ACt
On/off Activation/deactivation of low-temperature protection
tES
On
Test of low-temperature protection
SEt
[° C]
Setting of the cut-off temperature for the low-temperature protection
PrE
[...]
Read-out of the real-time temperature of the F-sensor
ALL
Activation of escorted alarm limits. [FAS]=locked limits/[ESC]=follows setpoint
dEF
Number of defrosts per 24 hours (4 is factory setting)
dPS
Reference sensor for the display (A, E or F)
Temperature settings
Key pad lock is
activated
or
.
for 3 seconds.