Electric low temperature protection
FP – Activation/deactivation of low temperature protection
Press and hold
+
Press
several times until "FP" is shown in the display
Press
to select "FP". "Act" is now shown in the display
Press
to select "Act"
Press
or
to activate/deactivate [1 = activated/ 0 = deactivated]
Press
to confirm the set value
Leave the user menu by pressing
FP – Setpoint of low temperature protection
Press and hold
+
Press
several times until "FP" is shown in the display
Press
to select "FP". "Act" is now shown in the display
Press
several times until "SEt" is shown in the display
Press
to select "SEt"
Press
or
to select the setpoint temperature for the low temperature protection
Press
to confirm the set value
Leave the user menu by pressing
the cabinet temperature is shown in the display
FP – Test of low temperature protection
Press and hold
+
Press
several times until "FP" is shown in the display
Press
to select "FP". "Act" is now shown in the display
Press
to progress to "tES"
Press
to select "tES" – test will then be performed
Leave the user menu by pressing
FP – Temperature of low temperature protection sensor
Press and hold
+
Press
several times until "FP" is shown in the display
Press
to select "FP". "Act" is now shown in the display
Press
several times until "Pre" is shown in the display
Press
to select "Pre"
Press
to show the low temperature protection sensor temperature
Leave the user menu by pressing
for more than 3 seconds
several times until the cabinet temperature is shown in the display
for more than 3 seconds
several times until
for more than 3 seconds
several times until the cabinet temperature is shown in the display
for more than 3 seconds
several times until the cabinet temperature is shown in the display