Local alarm settings
LHd – Setting the delay of the local upper alarm limit [min.]
Press and hold
+
Press
to select "LAL". "LHL" is now shown in the display
Press
several times until "LHd" is shown in the display
Press
to select "LHd". The delay of the upper alarm limit is now shown in the display
Press
or
to set the desired value for the delay of the upper alarm limit
Press
to confirm the set value
– The delay of the upper alarm limit is now set, proceed to other parameters by pressing
Leave the user menu by pressing
LLd – Setting the delay of the local lower alarm limit [min.]
Press and hold
+
Press
to select "LAL". "LHL" is now shown in the display
Press
several times until "LLd" is shown in the display
Press
to select "LLd". The delay of the lower alarm limit is now shown in the display
Press
or
to set the desired value for the delay of the lower alarm limit
Press
to confirm the set value
– The delay of the lower alarm limit is now set, proceed to other parameters by pressing
Leave the user menu by pressing
ATTENTION –
High and low temperature alarms set
up in the cabinets' controller (including
EAL alarms) must be accompanied by
additional redundant independent
external alarms to ensure maximum item
safety.
for more than 3 seconds
several times until the cabinet temperature is shown in the display
for more than 3 seconds
several times until the cabinet temperature is shown in the display