Page 1
Instructions for use – BioCompact II Models: 210/310/410/610 Instructions for use – BioCompact II Original instructions for use Original Instructions for use Item nr. 765042589 Revision nr. 20220804 English...
Quick guide – BioCompact II Parameter settings Temperature settings On/Off Defrosting Alarms A2, A3, A4, A5 Key pad lock is activated On/Off Press the button to turn the cabinet on. Press the button for 6 seconds to turn the cabinet off. The software version of the cabinet will be shown when turning the cabinet on, followed by the software variant.
Page 3
b i o l i n e Example: Setting the upper limits for the alarms; LHL Press and hold until the display shows LAL Press to select LAL, LHL is now shown in the display Press to select LHL, 25 is shown in the display Press to set the desired value for the upper temperature limit Press...
This instructions for use should be considered an integral part of the cabinet, and should be stored close to the cabinet and be easy to access. If the instructions for use is lost, please refer to your local distributor or Gram Commercial for a replacement.
Cabinet components BioCompact II 210, 310, 410, 210/210, 310/210 This part describes the main components pertinent to the user. ATTENTION – In the event of technical difficulties or breakdown, always contact Gram BioLine technical support or a Gram BioLine authorized service partner.
Page 7
14 . Digital display for controller Use the display to show the cabinet temperature and, to set the parameters described in this manual. ATTENTION – If parts show signs of damage; do not use cabinet and contact Gram BioLine or supplier for further aid.
BioCompact II 610 This part describes the main components pertinent to the user. ATTENTION – In the event of technical difficulties or breakdown, always contact Gram BioLine technical support or a Gram BioLine authorized service partner.
Page 9
13 . Digital display for controller Use the display to show the cabinet temperature and, to set the parameters described in this manual. ATTENTION – If parts show signs of damage; do not use cabinet and contact Gram BioLine or supplier for further aid.
To ensure efficient operation, the cabinet should not be installed in direct sunlight or close to heat sources. BioCompact II 210, 310, 410, 210/210, 310/210 RR with solid door +10ºC +35ºC The cabinet interior must not be exposed to corrosive atmospheres.
Page 11
b i o l i n e b i o l i n e Cabinets equipped with legs should be levelled as shown in the illustration below. For cabinets equipped with castors, the floor must be level to ensure stable positioning and safe use. When the cabinet is positioned, the 2 front castors should be locked.
Anti tilt bracket Cabinets with drawers and/or glass door must be secured to a stable vertical surface, ensuring that the cabinet can not tip over when the drawers are drawn to the outermost position, or the door is open. Brackets for securing is included. Find the instructions for the anti-tilt bracket below.
Page 13
The anti-tilt brackets must be fitted when installing the cabinet, ensuring that the users, surroundings and stored items are not damaged by a tilting cabinet. Find instructions on wall mounting of a BioCompact II 210 below, the same procedure applies for mounting 310, 210/210, 310/210 and 410.
Surroundings The cabinet must be installed in accordance with the illustrations below. 15-75 mm ≥ 15 mm ≥ 30 mm...
Page 15
b i o l i n e Do not cover the upper part of the cabinet. Do not use electrical appliances inside the cabinet. The cabinet is not suited for storing items that emit vapours, as this might lead to deterioration of cabinet performance and/or longevity.
Find instructions on setting external alarms in the Parameter settings section. Access to the voltage free contact is done according to the descriptions below: BioCompact II 210, 310, 210/210, 310/210, 410 Unscrew the preload cover on the back of the cabient to access the voltage free contact block.
Page 17
Placement of voltage free contact BioCompact II 610 BioCompact II 210, 310, 210/210, 310/210, 410 ly Open (NO) Common Common Common ly Open (NC) Common Normally closed circuit (NC) Common Common...
Never use the cabinet if the plug is damaged. The cabinet should be examined by a Gram Commercial service The cabinet is intended for connection to alternating current. The technician in such cases.
Page 19
BioCompact II 610 BioCompact II 210, 310, 210/210, 310/210, 410...
Equipotential bonding Exclusively for the models – BioCompact II 210, 310, 410, 210/210, 310/210 For installation in ATEX Cat. 3 Zone 2 areas, it is mandatory to have a equipotential bonding, it is not sufficient to use protective earth through the mains connection.
Start-up b i o l i n e The digital display Parameter settings Temperature settings On/Off Defrosting Alarms A2, A3, A4, A5 Key pad lock is activated On/Off • Parameter setting Gives access to the cabinets configurable parameters, such as Press to turn the cabinet on.
Walkthrough of menu Menu Access Local alarm settings [° C] Upper alarm limit. Code for activated alarm [A2] [° C] Lower alarm limit. Code for activated alarm [A3] [Min.] Delay of upper alarm limit [Min.] Delay of lower alarm limit On/off Door alarm.
Error codes b i o l i n e Display code Explanation - 0 - Door is open Door alarm “dAd” from LAL and/or EAL has been activated Local upper alarm LHL is or has been activated Local lower alarm LLL is or has been activated External upper alarm EHL is or has been activated External lower alarm ELL is or has been activated Error on the main cabinet sensor.
Local alarm settings Local high alarm Local low alarm LHL – Setting the upper alarm limit [° C] Press and hold for more than 3 seconds Press to select “LAL”. “LHL” is now shown in the display Press to select “LHL”. The upper alarm limit is now shown in the display Press to set the desired value for the upper alarm limit Press...
b i o l i n e b i o l i n e Local alarm settings LHd – Setting the delay of the local upper alarm limit [min.] Press and hold for more than 3 seconds Press to select “LAL”. “LHL” is now shown in the display Press several times until “LHd”...
On/Off local door alarm Delay for local door alarm dA – Activate/deactivate of local door alarm Press and hold for more than 3 seconds Press to select “LAL”. “LHL” is now shown in the display Press several times until “dA” is shown in the display Press to select “dA”.
b i o l i n e b i o l i n e Local acoustic settings bU – Activation/deactivation of the acoustic local alarms Press and hold for more than 3 seconds Press to select “LAL”. “LHL” is now shown in the display Press several times until “bU”...
External alarm settings External high alarm External low alarm EHL – Setting the external upper alarm limit [° C] Press and hold for more than 3 seconds Press several times until “EAL” is shown in the display Press to select “EAL”. “EHL” is now shown in the display Press to select “EHL”.
b i o l i n e b i o l i n e External high alarm delay External low alarm delay EHd – Setting the delay of the external upper alarm limit [min.] Press and hold for more than 3 seconds Press to proceed to “EAL”...
On/Off external door External door alarm delay alarm dA – Activation/deactivation of external door alarm Press and hold for more than 3 seconds Press to proceed to “EAL” Press to select “EAL”. “EHL” is now shown in the display Press several times until “dA”...
b i o l i n e External acoustic settings bU – Activation/deactivation of the acoustic external alarms Press and hold for more than 3 seconds Press to proceed to “EAL” Press to select “EAL”. “EHL” is now shown in the display Press several times until “bU”...
Parameter settings Sensor offset The temperature sensors connected to MPC controller can be offset independently of each other in the parameter cAL. Offset is used in cases where there are deviations in the cabinets actual operation compared to the display and/or control measurements by independent temperature monitoring.
Page 35
b i o l i n e Offset of the A-sensor Press and hold for more than 3 seconds Press several times until “cAL” is shown in the display Press to select “cAL”. “cA” is shown in the display Press to select “cA”...
Escorted/set alarm limits ALL – Setting of escorted/set alarm limits Press and hold for more than 3 seconds Press several times until “ALL” is shown in the display Press to select “ALL” Press to select set or escorted alarm limits Press to confirm the set value Leave the user menu by pressing...
b i o l i n e Defrosts/24 hours dEF – Number of defrosts Press and hold for more than 3 seconds Press several times until “dEF” is shown in the display Press to select “dEF” Press to set the desired amount of defrosts per 24 hours (factory setting is 4) Press to confirm the set value Leave the user menu by pressing...
Display sensor Press and hold for more than 3 seconds Press several times until “dPS” is shown in the display Press to select “dPS” Press to select either the A- or E-sensor Press to confirm the set value Leave the user menu by pressing several times until the cabinet temperature is shown in the display Please note: that dPS only changes the reference sensor for the display, and not the reference sensor for the alarms.
b i o l i n e Electric low temperature protection FP – Activation/deactivation of low temperature protection Press and hold for more than 3 seconds Press several times until “FP” is shown in the display Press to select “FP”. “Act” is now shown in the display Press to select “Act”...
The cabinet is not suited for storing items that emit vapours, as they might corrode the cabinet and its components. BioCompact II 410 The cabinet interior must not be exposed to corrosive atmospheres. BioCompact II 610...
Cleaning agents containing chlorine or compounds of chlorine as well as other corrosive agents, may not be used, as they may cause corrosion. The location of the condensers for both bottom- and top-mounted compressors is illustrated below. BioCompact II 610 BioCompact II 210, 310, 410, 210/210...
The door gasket should be cleaned regularly with a mild soap solution. If a gasket is to be replaced, please contact your local Gram BioLine distributor. The illustration below shows the location of the door gasket on a BioCompact II 410. Door gasket...
Defective parts must be replaced with original parts from Gram BioLine. Gram BioLine can only guarantee functional and safety requirements on the cabinets, if above mentioned is adhered to.
BioCompact II 210, 310, 210/210, 310/210, and 410 Defrost water is led through a tube in the insulation to a re-evaporation tray in the compressor room in the cabinet. Below the re-evaporation trays for a BioCompact II 410 and BioCompact II 610 are shown. BioCompact II 410...
Access port All BioCompact ll cabinets are equipped with an access port on the back of the cabinets, this can be used to easily fit external sensors. All access ports are constructed in the same fashion, with a conical polystyrene plug (fitted from the back of the cabinet) and a plastic cap (fitted from the inside of the cabinet).
b i o l i n e Important IMPORTANT! There may be sharp edges on the cabinet housing, compressor room, and interior, show due diligence when handling the cabinet, neglect of these precautions can lead to injuries. Danger of wedging of body parts in the frame slot between the door and the cabinet, show due diligence when opening and closing the cabinet door.
Disposal Electrical and electronic equipment (EEE) contains materials, components and substances that can be dangerous and harmful to human health and the environment if the waste (WEEE) is not disposed of properly. Contact you local BioLine distributor when the cabinet needs to be disposed of. Products that are labelled with a “crossed-out wheelie bin ‘is electric and electronic equipment.
IQ, OQ & PQ IQ, OQ and PQ documentation is readily available for all BioCompact ll cabinets . For Installation, Operation and Performance Qualification documentation please go to our website: www .gram-bioline .com...
BioCompact II 210, 310, 410, 210/210, 310/210 Technical specifications Data Connection 230 VAC, 50 Hz Control Unit Gram Control Unit with voltage free contact, alarms and offset function Alarms Acoustic and visual, High/Low temperature alarms and door alarm Alarm ports Voltage free contact Access port 1 pcs.
BioCompact II RR210H BioCompact II RR210H BioCompact II RR210H Solid door Glass door Technical specifications Data Technical specifications Data Temperature range +2/+20°C Temperature range +2/+20°C Ambient temperature range +10/+35°C Ambient temperature range +10/+32°C...
BioCompact II RR310H BioCompact II RR310H BioCompact II RR310H Solid door Glass door Technical specifications Data Technical specifications Data Temperature range +2/+20°C Temperature range +2/+20°C Ambient temperature range +10/+35°C Ambient temperature range +10/+32°C Software variant Software variant K-Value 0.35 W/(m2*K) K-Value 0.50 W/(m2*K)
BioCompact II RR410H BioCompact II RR410H BioCompact II RR410H Solid door Glass door Technical specifications Data Technical specifications Data Temperature range +2/+20°C Temperature range +2/+20°C Ambient temperature range +10/+35°C Ambient temperature range +10/+32°C...
BioCompact II RF210H BioCompact II RF310H BioCompact II RF210H BioCompact II RF310H Solid door Solid door Technical specifications Data Technical specifications Data Temperature range -25/-5°C Temperature range -25/-5°C Ambient temperature range +10/+35°C Ambient temperature range +10/+35°C Software variant Software variant K-Value 0.35 W/(m2*K)
BioCompact II RF410H BioCompact II RF410H Solid door Technical specifications Data Temperature range -25/-5°C Ambient temperature range +10/+35°C Software variant K-Value 0.35 W/(m2*K) ATEX marking II 3G Ex ec nC ic llB T3 Gc...
BioCompact II RR210/RR210H BioCompact II RR210/RR210H BioCompact II RR210/RR210H Solid door Glass door Technical specifications Data Technical specifications Data Temperature range +2/+20°C Temperature range +2/+20°C Ambient temperature range +10/+35°C Ambient temperature range +10/+32°C Software variant Software variant K-Value 0.35 W/(m2*K);...
BioCompact II RR210/RF210H BioCompact II RR210/RF210H BioCompact II RR210/RF210H Solid door Glass door (RR) Technical specifications Data Technical specifications Data Temperature range +2/+20°C and -25/-5°C Temperature range +2/+20°C and -25/-5°C Ambient temperature range +10/+35°C...
BioCompact II RF210/RF210H BioCompact II RF210/RF210H Solid door Technical specifications Data Temperature range -25/-5°C Ambient temperature range +10/+35°C Software variant K-Value 0.35 W/(m2*K) ATEX marking II 3G Ex ec nC ic llB T3 Gc ATEX certificate DTI 21ATEX0160X Gross volume...
BioCompact II RR310/RF210H BioCompact II RR310/RF210H BioCompact II RR310/RF210H Solid door Glass door (RR) Technical specifications Data Technical specifications Data Temperature range +2/+20°C and -25/-5°C Temperature range +2/+20°C and -25/-5°C Ambient temperature range +10/+35°C...
BioCompact II RR210G BioCompact II RR210G BioCompact II RR210G Solid door Glass door Technical specifications Data Technical specifications Data Temperature range +2/+20°C Temperature range +2/+20°C Ambient temperature range +10/+35°C Ambient temperature range +10/+32°C Software variant Software variant K-Value 0.35 W/(m2*K) K-Value 0.49 W/(m2*K)
BioCompact II RR310G BioCompact II RR310G BioCompact II RR310G Solid door Glass door Technical specifications Data Technical specifications Data Temperature range +2/+20°C Temperature range +2/+20°C Ambient temperature range +10/35°C Ambient temperature range +10/+32°C...
BioCompact II RR410G BioCompact II RR410G BioCompact II RR410G Solid door Glass door Technical specifications Data Technical specifications Data Temperature range +2/+20°C Temperature range +2/+20°C Ambient temperature range +10/+35°C Ambient temperature range +10/+32°C Software variant Software variant K-Value 0.35 W/(m2*K) K-Value 0.51 W/(m2*K)
BioCompact II RF210G BioCompact II RF310G BioCompact II RF210G BioCompact II RF310G Solid door Solid door Technical specifications Data Technical specifications Data Temperature range -25/-5°C Temperature range -25/-5°C...
BioCompact II RF410G BioCompact II RF410G Solid door Technical specifications Data Temperature range -25/-5°C Ambient temperature range +10/+35°C Software variant K-Value 0.35 W/(m2*K) ATEX marking II 3G Ex ec nC ic llB T3 Gc ATEX certificate DTI 21ATEX0160X Gross volume...
BioCompact II RR210/RR210G BioCompact II RR210/RR210G BioCompact II RR210/RR210G Solid door Glass door Technical specifications Data Technical specifications Data Temperature range +2/+20°C Temperature range +2/+20°C Ambient temperature range +10/+35°C Ambient temperature range +10/+32°C...
BioCompact II RR210/RF210G BioCompact II RR210/RF210G BioCompact II RR210/RF210G Solid door Glass door (RR) Technical specifications Data Technical specifications Data Temperature range +2/+20°C and -25/-5°C Temperature range +2/+20°C and -25/-5°C Ambient temperature range +10/+35°C Ambient temperature range +10/+32°C Software variant...
BioCompact II RF210/RF210G BioCompact II RF210/RF210G Solid door Technical specifications Data Temperature range -25/-5°C Ambient temperature range +10/+35°C Software variant K-Value 0.35 W/(m2*K) ATEX marking II 3G Ex nA nC nL llB T3 Gc...
BioCompact II RR310/RF210G BioCompact II RR310/RF210G BioCompact II RR310/RF210G Solid door Glass door (RR) Technical specifications Data Technical specifications Data Temperature range +2/+20°C and -25/-5°C Temperature range +2/+20°C and -25/-5°C Ambient temperature range +10/+35°C Ambient temperature range +10/+32°C Software variant...
BioCompact II 610 Technical specifications Data Connection 230 VAC, 50 Hz Control Unit Gram Control Unit with voltage free contact, alarms and offset function Alarms Acoustic and visual, High/Low temperature alarms and door alarm Alarm ports Voltage free contact Access port 1 pcs.
BioCompact II RF610H BioCompact II RR610H BioCompact II RR610H Solid door Glass door Technical specifications Data Technical specifications Data Temperature range +2/+20°C Temperature range +2/+20°C Ambient temperature range +10/+43°C Ambient temperature range +10/+38°C...
Page 72
BioCompact II RF610H Solid door Technical specifications Data Temperature range -25/-5°C Ambient temperature range +10/+43°C Software variant K-Value 0.31 W/(m2*K) ATEX marking II 3G Ex nA nC nL llB T2 Gc ATEX certificate – Gross volume 583 litres Net volume 419 litres Dimensions –...
BioCompact II RR610G BioCompact II RR610G BioCompact II RR610G Solid door Glass door Technical specifications Data Technical specifications Data Temperature range +2/+20°C Temperature range +2/+20°C Ambient temperature range +10/+43°C Ambient temperature range +10/+38°C...
BioCompact II RF610G BioCompact II RF610G Solid door Technical specifications Data Temperature range -25/-5°C Ambient temperature range +10/+43°C Software variant K-Value 0.31 W/(m2*K) ATEX marking II 3G Ex nA nC nL llB T2 Gc ATEX certificate – Gross volume 583 litres...
Declaration of conformity English EC Declaration of Conformity We, Gram Commercial declare under sole responsibility that the following products: Range: BioCompact II Model: 210, 310, 410, 210/210 & 310/210 Refrigeration: R600a & R134a Product description: Refrigerators and freezer for laboratory and biostorage...
Page 77
English EC Declaration of Conformity We, Gram Commercial declare under sole responsibility that the following products: Range: BioCompact II Model: Refrigeration: R600a, R290, R404A & R134a Product description: Refrigerators and freezer for laboratory and biostorage...
Page 78
English EC Declaration of Conformity We, Gram Commercial declare under sole responsibility that the following products: Range: BioCompact II (Accessorie code 69) Model: 210, 310, 410, 210/210 & 310/210 Refrigeration: R600a & R134a Product description: Refrigerators and freezer for laboratory and biostorage...
Page 79
English EC Declaration of Conformity We, Gram Commercial declare under sole responsibility that the following products: Range: BioCompact II (Accessorie code 69) Model: Refrigeration: R600a, R290, R404A & R134a Product description: Refrigerators and freezer for laboratory and biostorage...
Aage Grams Vej 1 6500 Vojens All rights to this drawing belongs to Gram Commercial A/S, cf. law of Copyright. Thus the drawing should notwithout our written permission be copied, presented or passed over to a third person. Misuse will lead to prosecution.
Need help?
Do you have a question about the BioCompact II and is the answer not in the manual?
Questions and answers