Local Acoustic Settings - Gram BioCompact II Instructions For Use Manual

Table of Contents

Advertisement

Local acoustic settings

bU – Activation/deactivation of the acoustic local alarms
Press and hold
+
Press
to select "LAL". "LHL" is now shown in the display
Press
several times until "bU" is shown in the display
Press
to select "bU".
Press
or
to activate/deactivate the local acoustic alarms [1 = activated/0 = deactivated]
Press
to confirm the set value
– The local acoustic alarms is configured, proceed to other parameters by pressing
Leave the user menu by pressing
ATTENTION –
High and low temperature alarms set
up in the cabinets' controller (including
EAL alarms) must be accompanied by
additional redundant independent
external alarms to ensure maximum item
safety.
for more than 3 seconds
several times until the cabinet temperature is shown in the display
b i o l i n e
b i o l i n e
, and then navigate by using
or
.
29

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the BioCompact II and is the answer not in the manual?

This manual is also suitable for:

210310410610

Table of Contents