External alarm settings
External high alarm
External low alarm
EHL – Setting the external upper alarm limit [° C]
Press and hold
+
Press
several times until "EAL" is shown in the display
Press
to select "EAL". "EHL" is now shown in the display
Press
to select "EHL". The external upper alarm limit is now shown in the display
Press
or
to set the desired value for the external upper alarm limit
Press
to confirm the set value
– The external upper alarm limit is now set, proceed to other parameters by pressing
Leave the user menu by pressing
ELL – Setting the external lower alarm limit [° C]
Press and hold
+
Press
several times until "EAL" is shown in the display
Press
to select "EAL". "EHL" is now shown in the display
Press
to proceed to "ELL"
Press
to select "ELL". The external lower alarm limit is now shown in the display
Press
or
to set the desired value for the external lower alarm limit
Press
to confirm the set value
– The external lower alarm limit is now set, proceed to other parameters by pressing
Leave the user menu by pressing
ATTENTION –
High and low temperature alarms set
up in the cabinets' controller (including
EAL alarms) must be accompanied by
additional redundant independent
external alarms to ensure maximum item
safety.
30
30
for more than 3 seconds
several times until the cabinet temperature is shown in the display
for more than 3 seconds
several times until the cabinet temperature is shown in the display
, and then navigate by using
, and then navigate by using
or
.
or
.