Error codes

Display code
Explanation
- 0 -
Door is open
A1
Door alarm "dAd" from LAL and/or EAL has been activated
A2
Local upper alarm LHL is or has been activated
A3
Local lower alarm LLL is or has been activated
A4
External upper alarm EHL is or has been activated
A5
External lower alarm ELL is or has been activated
Error on the main cabinet sensor. The refrigeration system will use an emergency program to make the cabi-
F1
net run. Temperature stability will be affected. Service is required
F2
Error on the evaporator sensor. Service is required
F3
Error on the 1. condenser sensor. Service is required
F5
Error on the extra sensor. Service is required
F6
Error on low temperature protection sensor. Service is required
F7 indicates that the temperature of the condenser is too high. Turn off the cabinet and check that the
F7
condenser is not covered by undesirable items, and ensure that the condenser (and possibly filter) is clean.
Service is required if the problem is not alleviated
Acknowledge an acoustic alarm
Alarm code A1: Press
to acknowledge.
Temperature alarm codes A2 and/or A3: Flashes in the display. Press
The display will continue to flash if the temperature is outside the alarm limits.
Latching alarms: A2, A3, A4, A5
Due to the potential implications of alarms, the red danger triangle light will continue to remain on to warn the user of an alarm having
occured. Regardless of acknowleding, the light will remain on until an alarm history reset has been performed (holding
more than 3 seconds), note that this will also reset temperature and alarm history.
Reading the alarm history – Example [A2]
[A2] flashes in the display – The temperature has exceeded the set value for the upper temperature limit, LHL.
Press
to acknowledge the [A2]. The display continues to flash, indicating that there is information in the alarm history.
Press
, Htt (High temperature time) is shown, press
Press
to return to Htt. Press
Press
to return to Ht and press
The procedure for reading an [A3] alarm is identical, apart from entering the alarm history with
below set limits, the parameters are Ltt and Lt.
A flashing display with no alarm codes indicates that the alarm codes have been acknowledged, but the alarm history contains
information.
Resetting Max/Min and alarm history
Resetting of the Max/Min and alarm history is done by holding
when reset is complete.
to see for how long the temperature was above the set alarm limit.
to reach Ht (Highest temperature). Press
again to leave the alarm history.
to acknowledge.
to read the highest recorded temperature during Htt.
and
for more than 3 seconds, an acoustic signal will be given
b i o l i n e
and
. When reading out temperatures
for
25