9 CONTROL PANEL
Patient's
right/left side
3 angles lateral, left
1 cm
First exposure = Selected angle - 7°
Second exposure = Angle selected by the user
(factory default angle: 17°)
Third exposure = Selected angle + 7°
24
Planmeca ProOne
If the patient's right and left side are asymmetric you will
need to set the target position and imaging angle (lateral
Symmetric/
views) separately on the other side. To do this, touch the
asymmetric
icon for symmetric/asymmetric setting to select the
setting
asymmetric setting and adjust the other side as required.
In the asymmetric setting the icon shows a divided jaw
(two lines in the middle).
NOTE
In the asymmetric setting, touching the target position
arrows on the main TMJ display will adjust the imaging
position only on the patient's left side.
The right/left buttons at the top of the display allow you to
select on which side the exposures will be taken. You can
set the unit to take TMJ exposures on one side or on both
sides. The button(s) of the selected side(s) will be
highlighted. Note that the buttons refer to the patient's
(not your) right and left side.
NOTE
In multi-angle exposures the selected imaging angle is in
image no. 2.
1 2 3
3.
2.
Imaging angle selected by the user
1.
User's Manual