9 Diagnosis and service

40
9 Diagnosis and service

9.1 Maintenance

When the device is used correctly, no maintenance is required in
normal operation.
The corresponding source container must be checked in regular
intervals. You can find further information in the operating instructions
manual of the source container.

9.2 Status messages

The instrument has self-monitoring and diagnosis according to NE 107
and VDI/VDE 2650. Status messages are outputted on the device
status. Detailed messages are visible under diagnosis via DTM,
indicating and adjustment module (dependent on the sensor type) and
EDD; there also in the respective colour. The messages are divided
into the following categories:
1
Fig. 17: Pictograms of the status messages
Failure - red
1
2
Function check - orange
Out of specification - yellow
3
4
Maintenance - blue
Failure: Due to a malfunction in the instrument, a failure message is
outputted. This status message is always activated and cannot be
deactivated.
Function check: The instrument is in operation, the measured value is
temporarily invalid (for example during simulation). This status
message is always activated and cannot be deactivated by the user.
Out of specification: The measured value is unstable because the
instrument specification is exceeded (e.g. electronics temperature).
This status message is always inactive and can be activated via DTM
and EDD. The error codes and text messages are dispayed via the
indicating and adjustment module as well as DTM and EDD in the
control system. Additional information on error statstics is displayed in
the menu Diagnosis under "Device status" in the indicating and
adjustment module as well as in the DTM and EDD.