Vega POINTRAC 31 Operating Instructions Manual page 42

8/16 ma/hart - four-wire
Table of Contents

Advertisement

9 Diagnosis and service
42
Code
Cause
Text mes-
sage
F034
l Electronics defective
EPROM hard-
ware error
F035
l Error in the internal instru-
ment communication
EPROM data
error
F036
l Error during software up-
date
Faulty pro-
gram memory
F037
l Error in RAM
RAM hard-
ware error
F038
l Connection cable to the
Slave instrument interrup-
Slave signals
ted
failure
l Instrument not defined as
Slave instrument
F040
l Sensor defective
Hardware er-
ror
F041
l Error in the measured va-
lue recording
Photomultipli-
er error
F045
l Error on the current output l Check wiring of the current
Error in the
current output
F052
l Invalid parameter adjust-
ment
Faulty confi-
guration
F122
l Instrument addresse was
assigned several times
Double add-
ress on the
multisensor
communicati-
on bus
F123
l External instruments
cause radioactive radiation
X-ray alarm
Removal
l Exchange of the electro-
nics
l Carry out a reset
l Exchange of the electro-
nics
l Repeat software update
l Exchange of the electro-
nics
l Exchange of the electro-
nics
l Define instrument as Slave
l Check the connection cab-
le to the Slave instrument
l Exchange of the electro-
nics
l Exchange of the electro-
nics
output
l Exchange of the electro-
nics
l Carry out a reset
l Change instrument addr-
esses
l Determine reason for X-ray
alarm
l In case of a brief X-ray
alarm, shut down the in-
strument (switching) out-
puts for this time
POINTRAC 31 • 8/16 mA/HART - four-wire