Vega POINTRAC 31 Operating Instructions Manual page 43

8/16 ma/hart - four-wire
Table of Contents

Advertisement

Function check
Out of specification
POINTRAC 31 • 8/16 mA/HART - four-wire
Code
Cause
Text mes-
sage
F124
l Radiation dose too high
Alarm due to
increased ra-
diation
F125
l Ambient temperature on
the housing outside the
Ambient tem-
specification
perature too
high
The following table shows the error codes and text messages in the
status message "Function check" and provides information on causes
as well as corrective measures.
Code
Cause
Text mes-
sage
C029
l Simulation active
Simulation
The following table shows the error codes and text messages in the
status message "Out of specification" and provides information on
causes as well as corrective measures.
Code
Cause
Text mes-
sage
S017
l Accuracy outside the spe-
cification
Accuracy out-
side the spe-
cification
S025
l Bad linearization table
Bad lineariza-
tion table
S038
l Ambient temperature too
high/too low
Ambient tem-
perature too
high/too low
9 Diagnosis and service
Removal
l Determine reason for inc-
reased radiation
l Cool (heat) the instrument
or protect it with isolation
material against cold or
radiation heat
Removal
l Finish simulation
l Wait for the automatic end
after 60 mins.
Removal
l Correct adjustment data
l Carry out linearization
l Protect instrument with
isolating material against
extreme temperatures
43