Couplings and
connectors
Turbine air
filter
Mask
6-6
WARNING !!
It is necessary to have at least one spare patient circuit in
stock for routine use and /or accidental breaks.
Dismantle and clean: sterilize in an
autoclave, disinfect with steam or
chemically.
Check that there are no splits and replace
them if they are damaged.
Dismantle and clean with hot water and a
neutral detergent solution.
Check that there are no splits and replace
them if they are damaged.
Do not clean or re-use disposable circuit tubes.
Components that cannot be destroyed should be sterilized
and disinfected according to local standards.
Perform daily cleaning of the mask
following the instructions of the
responsible doctors or recommended
by the Manufacturer.
Hang up the clean mask to provide
that it is completely dry before use.
Always clean the mask and the hoses
or use a new mask in case the lung
ventilator must be used with a
different patient.
If the lung ventilator is used with more
than one patient in the clinic, insert an
antibacterial filter between the patient
outlet and the hose.
Risk of injury for the patient
Before using again, eliminate
any humidity inside the
components by means of
compressed air.
Before mounting the filter
again, eliminate any humidity
inside the components by
means of compressed air.
See Manufacturer's
instructions.
User manual, DU3104101