O
2
Concentration percentage delivered to the patient can be set
from 21% to 100%.
Pressure Low (cmH
O) - Pressure High (cmH
O)
2
2
Pressure levels to be set in APRV mode.
Pause (%)
Inspiratory pause time. The "inspiratory pause time" is displayed
on the screen in % (% of the inspiratory time). It is also used to
calculate the lung mechanics parameters (resistance and static
compliance).
PEEP (cmH
O)
2
Positive airway pressure value during expiratory phase.
Pinsp (cmH
O)
2
Maximum airway pressure limit value. The parameter is used in
pressure-controlled modes to fix an operating limit for the
airway pressure that shall not be exceeded.
PMax (cmH
O)
2
Maximum airway pressure limit.
Pmin (cmH
O)
2
Minimum airway pressure limit.
PS (cmH
O)
2
Positive airway support pressure value during inspiratory phase.
RR (bpm)
Lung ventilator respiratory rate.
RRsimv (bpm)
Value of forced respiratory rate in SIMV mode.
Sigh. Ampl. (%)
Sigh. Percentage increase of the set Vti.
2-22
User manual, DU3104101