A.6.3 Table 3
The device is intended for use in the electromagnetic environment specified below. The
customer or the user of the device should assure that it is used in such an environment.
Immunity test
EN 60601-1-2
RF conducted
EN 61000-4-6
RF Radiated
IEC 61000-4-3
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,
should be less than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with the following symbol:
Immunity aspects at R.F.
Test level
Compliance
level
3 Veff from
3 Veff from
150kHz to
150kHz to
80MHz
80MHz
3 Veff from
3 Veff from
80MHz to
80MHz to
2,5 GHz
2,5 GHz
Electromagnetic environment -
guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the device, including cables,
than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
d = 1,2 ⋅√P from 150kHz to 80MHz
d = 1,2 √P from 80 MHz to 800 MHz
d = 2,3 √P from 800 MHz to 2,5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m)