It shows the ratio between the inspiration time and the expiration
time.
It shows the oxygen concentration value (as percentage) inhaled by
the patient.
The inhaled oxygen concentration value is read by the system by
means of the oxygen cell installed on the inspiratory line.
It shows the current volume value during patient's expiratory phase:
the unit of measurement is ml.
The value is detected by the flow sensor installed on the expiratory
line.
It shows the volume value expired by the patient per minute: the unit
of measurement is L/min.
You can also calculate this value using the formula: current volume
(Vte) x respiratory frequency (RF).
It shows the value of end expiration CO
The value is detected by the gas sensor analyzer installed on the
expiratory line.
Description
Vte: respiratory parameter
ml: unit of measurement
503: value set by the User
1000 - 100: alarm limits
(end-tidal CO
).
2
2
2-25