When searching for or reading in coordinate data or searching for code data,
the search stopped either because the item in question does not exist or the
data volume is large.
No solution
The calculation of the instrument station coordinates during resection does
not converge.
Assess the results and if necessary, perform the observation again.
Intersect point could not be calculated. Either necessary data items were not
input or the Intersect point does not exist.
North/East is null, Read error
The Northing or Easting field of the given coordinate is null.
Input the coordinate.
Out of range
The tilt of the instrument exceeds the tilt angle compensation range during
measurement.
Level the instrument again.
"7.2 Levelling"
A direction which does not cross with the base plane during plane offset
measurement.
Out of value
During gradient % display, the display range (less than ±1000%) has been
exceeded.
During REM measurement, either the vertical angle has exceeded horizontal
±89° or the measured distance is greater than 9999.999m.
Install the instrument station far from the target.
The instrument station coordinates calculated during resection are too high.
Perform the observation again.
During setting-out line measurement, scale factor has been less than
0.100000 or exceeded 9.999999.
During area calculation, results exceeded the display range.
32. WARNING AND ERROR MESSAGES