'Stutter' Laydown:
Laydown made up of
several larger drops.
May have gaps between
drops.
Uneven Laydown:
Middle strip varies in
thickness.
Shandon ClearVue™ Operator Guide
Possible Cause
Dispense Head
too high.
Pipes
Syringe
Air in
or
.
Dispense Head
too high.
Pipes
Syringe
Air in
or
.
Solution
Dispense Head
Check
fully screwed down.
Contact Thermo Fisher
Scientific Service
Department.
Flush
system until all bubbles
have been expelled.
Dispense Head
Check
fully screwed down.
Contact Thermo Fisher
Scientific Service
Department.
Purge
system until all bubbles
have been expelled.
A79210100 - Iss 2
is
is
75