5-2 - User Cleaning
and Maintenance
5-2-1 - Changing
Seals
90
A79210100 - Iss 2
The following sections give details on how to carry out basic
cleaning and maintenance procedures.
If a problem occurs which is not covered in these sections contact
Thermo Fisher Scientific Service Department.
O-Rings
Rubber
are used to ensure air-tight or water-tight seals
either for the reliable operation of the instrument, or for the safety
of the user.
If these seals become encrusted with mountant their ability to
create a good seal is impaired and they must be replaced.
There are two user serviceable seals on the Shandon ClearVue.
These are located on:
Mountant Bottle
•
The
Xylene Tray
•
The
O-Ring Seal
To remove the
replacement:
1.
Ensure that there are no
being processed within the Shandon ClearVue.
2.
Unscrew and remove the lid of the
Bottle
.
Shandon ClearVue™ Operator Guide
spout.
.
Mountant Bottle
on the
Baskets
currently
Mountant
spout for
O-Ring
Seal
O-Ring
Seal