Otto Bock C-Leg Prosthetic System Instructions For Use Manual page 18

Table of Contents

Advertisement

Switching between 1st and 2nd modes without the remote control:
3 × in 1
second
1 Bounce up and down on the forefoot at least 3 times in one second while maintaining continuous ground con-
tact. The foot must bear at least 70% of the maximum load. When relieving the foot, the foot must bear at least
15% of the maximum load. You will hear a beep signal.
2 Lift the leg for at least one second and extend it to the rear (no ground contact).
3 The C-Leg confirms the switching and changes to the respective other mode:
• Activation of the 2nd mode = 2 short beep signals (joint changes from 1st mode to 2nd mode)
• Activation of the 1st mode = 1 short beep signal (joint changes from 2nd mode to 1st mode)
Switching between 1st and 3rd modes without the remote control:
3 × in 1
second
1 While maintaining contact with the ground, bounce on the heel at least 3 × within a second. During this proc-
ess, at least 70 % of the maximum heel load must be placed on the foot. When the load is reduced, the minimum
value must not be less than 15 % of the maximum heel load. A beep signal sounds.
2 Take weight off the leg for at least one second.
3 The C-Leg confirms the switching process and switches to the other mode:
• Activating 3rd mode = 3 short beeps (joint switches from 1st mode into 3rd mode)
• Activating 1st mode = 1 short beep (joint switches from 3rd mode into 1st mode)
18 Ottobock
1
1 second
1
1
1
2
2
2
1 second
2
C-Leg Prosthetic System
3
3

Advertisement

Table of Contents
loading

Table of Contents