5 Preparing the product for use
CAUTION
Incorrect alignment, assembly or adjustment
Injury due to incorrectly installed or adjusted as well as damaged prosthet­
ic components
► Observe the alignment, assembly and adjustment instructions.
NOTICE
Grinding the prosthetic foot or footshell
Premature wear resulting from damage to the product
► Do not grind the prosthetic foot or footshell.
5.1 Alignment
INFORMATION
There is a plastic protector on the product's pyramid. It protects the con­
necting section from scratches during the alignment and fitting of the pros­
thesis.
► Remove the protector before the patient leaves the workshop/fitting
area.
5.1.1 Applying/removing the footshell
INFORMATION
► Pull the protective sock over the prosthetic foot to prevent noises in the
footshell.
► Always use the prosthetic foot with the footshell.
► Apply or remove the footshell as described in the footshell instructions
for use.
5.1.2 Bench Alignment
TT bench alignment
Required materials: 662M4 goniometer, 743S12 heel height measuring device,
743A80 50:50 gauge, alignment tool (e.g. 743L200 L.A.S.A.R. Assembly or 743A200
PROS.A. Assembly)
Assemble and align the prosthetic components in the alignment apparatus according
to the following specifications:
16
Bench alignment process